GUI reference
R&S
®
ZNB/ZNBT
798User Manual 1173.9163.02 ─ 62
Select some matrices and tap "OK" to remove them from the RF connection configura-
tion or tap "Cancel" to quit the dialog without removing a switch matrix from the RF
connection configuration.
Delete All
Opens a confirmation dialog
Either select "OK" to remove all switch matrices from the RF connection configuration,
or "Cancel" to quit the dialog without removing a switch matrix.
Remote command:
SYSTem:COMMunicate:RDEVice:SMATrix:DELete
Renumber Test Ports
Use this function at any time to generate the default test port assignment for the given
matrix VNA port connections.
The following RF ports are used as test ports (and numbered from 1 to N):
All matrix test ports that can be switched to at least one of the connected matrix
VNA ports
All VNA ports that are not connected to a switch matrix
Note that "Renumber Test Ports" does not preserve manually assigned test port num-
bers or "Unused" flags. See "Edit Test Port Connection" on page 799.
Remote command:
n.a.
Edit Matrix VNA Port Connections
To edit a matrix VNA port connection, tap on the corresponding port symbol.
Setup softtool

Table of Contents