Command reference
R&S
®
ZNB/ZNBT
872User Manual 1173.9163.02 ─ 62
Example:
CALCulate1:EYE:MASK:DATA?
returns something like
'Eye Mask
------------------------------
Fail Condition Type Samples
Violation Tolerance 1
Total Number of Samples 10342
Mask 1 (Top) Active
Samples Hits 366
Fail Rate 3.539 %
Test Result Fail
Mask 2 (Bottom) Not Active
Samples Hits ------
Fail Rate ------
Test Result ------
Mask 3 (Octagon) Not Active
Samples Hits ------
Fail Rate ------
Test Result ------
'
Usage:
Query only
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Mask Test On" on page 703
CALCulate<Chn>:EYE:MASK:FAIL?
Returns 'Pass' or 'Fail' to indicate the result of the limit check in the related eye
diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram, if the eye test is not enabled (CALCulate<Chn>:EYE:MASK:
STATe ON), or if the mask is empty (see CALCu
late<Chn>:EYE:MASK:SHAPe:BOTTom|TOP|POLYgon:STATe.
Tip
Use CALCulate:CLIMits:FAIL? to perform a composite (global) limit check
The result is automatically recalculated whenever a relevant setting is changed, i.e.
a subsequent query will return the updated mask test result
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI command reference