Command reference
R&S
®
ZNB/ZNBT
877User Manual 1173.9163.02 ─ 62
Parameters:
<Boolean> *RST: OFF
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Mask Test On" on page 703
CALCulate<Chn>:EYE:MASK:VIOLation:RATE <ViolationRate>
Defines the violation rate (i.e. the share of bad samples) for the mask test in the related
eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
This rate will only be used if CALCulate<Chn>:EYE:MASK:FAIL:CONDition is set
to RATE.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<ViolationRate> Default unit: percent
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Test Settings" on page 706
CALCulate<Chn>:EYE:MASK:VIOLation:TOLerance <ViolationTolerance>
Defines the violation tolerance (i.e. the number of bad samples) for the mask test in the
related eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
This tolerance will only be used if CALCulate<Chn>:EYE:MASK:FAIL:CONDition
is set to SAMPles.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<ViolationTolerance>
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Test Settings" on page 706
CALCulate<Chn>:EYE:MEASurement:DATA?
Returns the measurement results of the related eye diagram (see "Eye Diagram
Results" on page 231).
SCPI command reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals