Command reference
R&S
®
ZNB/ZNBT
1119User Manual 1173.9163.02 ─ 62
<ImaginaryPart>
Example:
*RST; :DISP:WIND:TRAC:X:OFFS 1MHZ; :DISP:WIND:
TRAC:Y:OFFS 10
Create the default trace and shift it horizontally by 1 MHz, verti-
cally by 10 dB.
DISP:WIND:TRAC:Y:OFFS?
Query all response offset values. The response is 10,0,0,0.
Manual operation: See "Mag / Phase / Real / Imag" on page 363
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y[:SCALe]:AUTO <Activate>[,
<TraceName>]
Displays the entire trace in the diagram area, leaving an appropriate display margin.
The trace can be referenced either by its number <WndTr> or by its name
<TraceName>.
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
DISPlay[:WINDow<Wnd>][:STATe] ON). This suffix is
ignored if the optional <TraceName> parameter is used.
<WndTr> Existing trace number, assigned by means of DISPlay[:
WINDow<Wnd>]:TRACe<WndTr>:FEED. This suffix is ignored if
the optional <TraceName> parameter is used.
Setting parameters:
<Activate> ONCE
Activate the autoscale function.
<TraceName> Optional string parameter for the trace name, e.g. 'Trc4'. If this
optional parameter is present, both numeric suffixes are ignored
(trace names must be unique across different channels and win-
dows).
Example:
*RST; DISP:WIND:TRAC:Y:PDIV?; RLEV?
Query the value between two grid lines and the reference value
for the default trace. The response is 10;0.
DISP:WIND:TRAC:Y:AUTO ONCE; PDIV?; RLEV?
or:
DISP:WIND:TRAC:Y:AUTO ONCE, 'Trc1'; PDIV?;
RLEV?
Autoscale the default trace and query the scaling parameters
again. In general both values have changed.
Usage: Setting only
Manual operation: See "Auto Scale Trace" on page 329
SCPI command reference