SYSTem:TIME <Hours>, <Minutes>, <Seconds>
The command queries or defines the instrument's current time setting.
The setting command requires administrator rights; refer to "User accounts and pass-
word protection" on page 30.
Parameters:
<Hours> Range: 0...23
<Minutes> Range: 0...59
<Seconds> Range: 0...59
Example:
SYST:TIME?
Response: 12,0,0 - it is precisely 12 pm.
SYSTem:TRESet[:STATe]
<Boolean>
Defines the behavior of the INITiate<Ch>[:IMMediate][:DUMMy] and
INITiate[:IMMediate]:ALL.
Parameters:
<Boolean> OFF (0)
Previous trace data are preserved.
ON (1)
Previous trace data are deleted.
This is a global setting that is not affected by an instrument
reset. Factory default is OFF (0).
Manual operation: See "Restart: Set all Traces to 0" on page 756
SYSTem:TSLock <Type>
Locks the touchscreen functionality of the R&S ZNB/ZNBT, e.g. in order to prevent
inadvertent entries during remote control.
Parameters:
<Type> OFF | DIAGrams | SCReen
OFF – touchscreen (e.g. the remote screen) remains active. You
can use the buttons in "Remote" softtool panels. If you switch
back to manual control, the drag and drop functions (e.g. for
markers) are still available.
DIAGrams – lock the drag and drop functions in diagrams .
SCReen – lock the entire screen including the remote screen
buttons.
*RST: n/a (*RST does not affect the tochscreen lock set-
ting).
Example:
SYSTem:TSLock SCReen
Lock the entire screen.
Manual operation: See "Enabled / Lock Diagrams / Lock Screen" on page 749