Page #340 background imageLoading...
Page #340 background image
GUI reference
R&S
®
ZNB/ZNBT
340User Manual 1173.9163.02 ─ 62
Data to <Destination>
Stores the current state of the active data trace to the Destination memory trace. No
trace functions are applied to the stored trace.
Tips:
Use Data & Func to <Destination> to apply trace functions to the stored trace.
You can also create memory traces using the Import Complex Data dialog.
It is not possible to store Hold traces to memory.
For the relation between a data trace and its associated memory traces, see "Cou-
pling of data and memory traces" on page 339.
Remote command:
CALCulate<Chn>:MATH:MEMorize
TRACe:COPY
Data & Func to <Destination>
Stores the current state of the active data trace – including trace functions – to the
Destination memory trace.
Trace functions
The trace functions comprise the following mathematical operations:
Active trace math on unformatted and formatted traces, as configured in the Chap-
ter 5.5.4, "Math tab", on page 342.
A shift of the data trace (see "Shift Trace" on page 363).
Data to <Destination> stores the raw trace without the trace functions, "Data & Func to
<Destination>" stores the trace after it has been transformed using the trace functions.
For the relation between a data trace and its associated memory traces, see "Coupling
of data and memory traces" on page 339.
Remote command:
CALCulate<Chn>:MATH:MEMorize
TRACe:COPY:MATH
Destination
Selects the destination for the Data to <Destination> and Data & Func to <Destination>
operations.
The destination can be one of the following:
An existing memory trace of the active data trace.
The existing memory trace is overwritten.
"New Trace"
The data are copied to a new memory trace, associated to the current data trace.
Remote command:
n/a
Show <Active Data Trace>
Shows or hides the active data trace in the diagram.
Remote command:
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:SHOW
Trace Config softtool