GUI reference
R&S
®
ZNB/ZNBT
461User Manual 1173.9163.02 ─ 62
Ports
Selects the test ports to be calibrated.
Note: Calibration and port de-/activation.
The analyzer fimware automatically activates/deactivates ports during/after a (success-
ful) calibration:
Calibrated ports that were previously disabled, are automatically enabled as single-
ended logical ports.
An uncalibrated port that is not used by a measurement (i.e. the port is not required
by any trace of the related channel) is disabled.
Remote command:
The port parameters in many calibration commands define the calibrated ports.
Cal Unit
Displays the connected calibration units. The R&S ZNB/ZNBT auto-detects all calibra-
tion units which are connected to one of its USB ports. If several cal units are connec-
ted, one of them must be selected for calibration (active cal unit). A warning is dis-
played if the current sweep range of the network analyzer exceeds the characterized
frequency range of the calibration unit.
For background information, see Chapter 4.5.5, "Automatic calibration", on page 181.
Remote command:
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Characterization
See "Characterization" on page 451.
Temperature Compensation
Available for inline calibration systems (ICS) only. Applies to all inline calibration units
(ICUs) that are connected to the ICS and that are used with factory characterization
data.
If checked, the R&S ZNB/ZNBT firmware adjusts the factory characterization data
according to the ICU temperatures.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:TEMPcomp
Calibration Type / Source
Selects the calibration type for the selected physical ports. For an overview, refer to
Table 4-7.
The reflection calibration types can be used for any combination of physical ports:
reflection calibrations are performed for each selected port.
A transmission calibration type requires at least two physical ports. For the unidirec-
tional transmission calibration types ("Trans Norm", "One Path Two Ports"), the direc-
tion ("Source" port) must be specified in addition.
Cal softtool

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals