Command reference
R&S
®
ZNB/ZNBT
898User Manual 1173.9163.02 ─ 62
Setting parameters:
<String> Path to the Touchstone file, relative to the cache directory.
Usage: Setting only
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Clear All 1L Measurements/Clear Selected 1L Measure-
ments" on page 666
CALCulate:FMODel:DELT:M1L:DIFFmode <Boolean>
The one-length method (CALCulate:FMODel:DELT:METHod M1L) handles single-
ended and balanced measurement results differently. This command selects between
the two modes.
Parameters:
<Boolean> OFF (0)
Single-ended mode
ON (1)
Differential mode
*RST: OFF (0)
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Method" on page 662
CALCulate:FMODel:DELT:M1L:FILename <String>
CALCulate:FMODel:DELT:M2L:FILename <String>
CALCulate:FMODel:DELT:M3L:FILename <String>
For a Delta-L+ PCB characterization, this command loads the "1L Measurement" data
from a Touchstone file.
Parameters:
<String> Path to the Touchstone file, either absolute or relative to the
selected working directory (CALCulate:FMODel:DELT:
DIRectory).
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Load File" on page 665
CALCulate:FMODel:DELT:M1L:LENGth <Length>
CALCulate:FMODel:DELT:M2L:LENGth <Length>
CALCulate:FMODel:DELT:M3L:LENGth <Length>
Specifies the length of the 1st/2nd/3rd line in Delta-L measurements (CALCu
late:FMODel:DELT<Ph_pt>:M1L|M2L|M3L:MEASure).
Parameters:
<Length> Length in inches