CALCulate:FMODel:DELT:FREQuencies:USER:DELete:ALL
The Delta-L algorithm can calculate the loss-per-inch uncertainties either at default or
at custom frequencies (see CALCulate:FMODel:DELT:FREQuencies:
USEDefault).
This command clears the list of custom frequencies.
Usage: Event
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Frequencies" on page 663
CALCulate:FMODel:DELT:FREQuencies:USER:COUNt?
The Delta-L algorithm can calculate the loss-per-inch uncertainties either at default or
at custom frequencies (see CALCulate:FMODel:DELT:FREQuencies:
USEDefault).
This command returns the number of custom frequencies, defined using CALCulate:
FMODel:DELT:FREQuencies:USER:ADD etc.
Usage:
Query only
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Frequencies" on page 663
CALCulate:FMODel:DELT:M1L:CACHe:CLEar:ALL
If the 1-length method is used (CALCulate:FMODel:DELT:METHod M1L), the VNA
firmware builds up a cache of coupon measurements (Touchstone files) in the
DeltaL\1LMeasurementCache subfolder of the user data folder
C:\Users\Public\Documents\Rohde-Schwarz\Vna.
This command clears this cache.
Usage:
Event
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Clear All 1L Measurements/Clear Selected 1L Measure-
ments" on page 666
CALCulate:FMODel:DELT:M1L:CACHe:CLEar:SELected <String>
If the 1-length method is used (CALCulate:FMODel:DELT:METHod M1L), the VNA
firmware builds up a cache of coupon measurements (Touchstone files) in the
DeltaL\1LMeasurementCache subfolder of the user data folder
C:\Users\Public\Documents\Rohde-Schwarz\Vna.
This command deletes the selected file from this cache.
SCPI command reference