Command reference
R&S
®
ZNB/ZNBT
896User Manual 1173.9163.02 ─ 62
CALCulate:FMODel:DELT:FREQuencies:USEDefault <Boolean>
Delta-L computes the uncertainty in the loss-per-inch calculation at specified frequen-
cies.
This command specifies whether to use the tool's default frequencies or custom fre-
quencies, which can be defined using CALCulate:FMODel:DELT<Ph_pt>:FRE
Quencies:USER commands.
Parameters:
<Boolean> ON (1)
Use default frequencies
OFF (0)
Use custom frequencies
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Frequencies" on page 663
CALCulate:FMODel:DELT:FREQuencies:USER?
The Delta-L algorithm can calculate the loss-per-inch uncertainties either at default or
at custom frequencies (see CALCulate:FMODel:DELT:FREQuencies:
USEDefault).
This command returns the custom frequencies, defined using CALCulate:FMODel:
DELT:FREQuencies:USER:ADD etc.
Usage:
Query only
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Frequencies" on page 663
CALCulate:FMODel:DELT:FREQuencies:USER:ADD <UserFrequencies>
CALCulate:FMODel:DELT:FREQuencies:USER:DELete <Frequency>
The Delta-L algorithm can calculate the loss-per-inch uncertainties either at default or
at custom frequencies (see CALCulate:FMODel:DELT:FREQuencies:
USEDefault).
CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:USER:ADD|DELete adds/
deletes a frequency to|from the list of custom frequencies.
Setting parameters:
<Frequency> Custom frequency
Range: The frequency range of the analyzer
Default unit: Hz
Usage: Setting only
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Frequencies" on page 663