Parameters:
<String> Path to the Touchstone file, either absolute or relative to the
selected working directory (CALCulate:FMODel:DIRectory).
Options: R&S ZNB/ZNBT-K210
Manual operation: See "Load File" on page 632
CALCulate:FMODel:EZD:DUT:MEASure:FILename:CLEar
Invalidates the file path previously set using CALCulate:FMODel:EZD:DUT:
MEASure:FILename.
Usage:
Event
Options: R&S ZNB/ZNBT-K210
Manual operation: See "Load File / 1x Open Preset / 1x Short Preset" on page 630
CALCulate:FMODel:EZD<Ph_pt>:DUT[:STATe] <Boolean>
Defines the physical ports to which DUT + test fixture are connected.
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean> ON (1)
Port <Ph_pt> connected
OFF (0)
Port <Ph_pt> not connected
Options: R&S ZNB/ZNBT-K210
Manual operation: See "Active" on page 632
CALCulate:FMODel:EZD:GENerate:SIDE<1|2> <Boolean>
Tells the EZD tool to create fixture deembedding data at side 1, at side 2, or at both
sides.
Suffix:
<1|2>
.
Side
Parameters:
<Boolean> ON (1)
Create fixture deembedding data at the specified side (<1|2>)
OFF (0)
Do not create fixture deembedding data at the specified side
(<1|2>)
Options: R&S ZNB/ZNBT-K210
Manual operation: See "Generate Side 1/Generate Side2" on page 640