EasyManuals Logo

Agilent Technologies InfiniiVision 3000 X-Series Programmer's Guide

Agilent Technologies InfiniiVision 3000 X-Series
1282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1050 background imageLoading...
Page #1050 background image
1050 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
35 Obsolete and Discontinued Commands
:CHANnel:LABel
(see page 1138)
Command Syntax
:CHANnel:LABel <source_text><string>
<source_text> ::= {CHANnel1 | CHANnel2 | DIGital<d>}
<d> ::= 0 to (# digital channels - 1) in NR1 format
<string> ::= quoted ASCII string
The :CHANnel:LABel command sets the source text to the string that
follows. Setting a channel will also result in the name being added to the
label list.
Query Syntax
:CHANnel:LABel?
The :CHANnel:LABel? query returns the label associated with a particular
analog channel.
Return Format
<string><NL>
<string> ::= quoted ASCII string
NOTE
The :CHANnel:LABel command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :CHANnel<n>:LABel command (see page 269) or
:DIGital<n>:LABel command (see page 292).

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals