752 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
27 :SBUS<n> Commands
:SBUS<n>:SPI:TRIGger:TYPE
(see page 1138)
Command Syntax
:SBUS<n>:SPI:TRIGger:TYPE <value>
<value> ::= {MOSI | MISO}
The :SBUS<n>:SPI:TRIGger:TYPE command specifies whether the SPI
trigger will be on the MOSI data or the MISO data.
When triggering on MOSI data, the data value is specified by the
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA and
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:WIDTh commands.
When triggering on MISO data, the data value is specified by the
:SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA and
:SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh commands.
Query Syntax
:SBUS<n>:SPI:TRIGger:TYPE?
The :SBUS<n>:SPI:TRIGger:TYPE? query returns the current SPI trigger
type setting.
Return Format
<value><NL>
<value> ::= {MOSI | MISO}
See Also "Introduction to :TRIGger Commands" on page 877
":SBUS<n>:SPI:SOURce:MOSI" on page 747
":SBUS<n>:SPI:SOURce:MISO" on page 746
":SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA" on page 748
":SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA" on page 750
":SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh" on page 749
":SBUS<n>:SPI:TRIGger:PATTern:MOSI:WIDTh" on page 751
":TRIGger:MODE" on page 886