:MEASure Commands 20
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 423
:MEASure:DUAL:VRMS
(see page 1138)
Overview This measurement is available with the N2820A high sensitivity current
probe when both the Primary and Secondary probe cables are used. This
measurement joins the Zoom In waveform data below the probe's clamp
level with Zoom Out waveform data above the probe's clamp level to create
the waveform on which the measurement is made.
Command Syntax
:MEASure:DUAL:VRMS [<interval>][,<type>][,<source1>][,<source2>]
<interval> ::= {CYCLe | DISPlay}
<type> ::= {AC | DC}
<source1>,<source2> ::= CHANnel<n> with N2820A probe connected
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:DUAL:VRMS command installs a screen measurement and
starts an RMS value measurement.
The <interval> option lets you specify the measurement interval: either an
integral number of cycles, or the full screen. If <interval> is not specified,
DISPlay is implied.
The <type> option lets you choose between a DC RMS measurement and
an AC RMS measurement. If <type> is not specified, DC is implied.
If the optional source parameter(s) are specified, the currently specified
source(s) are modified.
Query Syntax
:MEASure:DUAL:VRMS? [<interval>][,<type>][,<source1>][,<source2>]
The :MEASure:DUAL:VRMS? query measures and outputs the RMS value
measurement.
Return Format
<value><NL>
<value> ::= calculated dc RMS value in NR3 format
See Also • ":MEASure:DUAL:CHARge" on page 418
• ":MEASure:DUAL:VAMPlitude" on page 419
• ":MEASure:DUAL:VAVerage" on page 420
• ":MEASure:DUAL:VBASe" on page 421
• ":MEASure:DUAL:VPP" on page 422
• "Introduction to :MEASure Commands" on page 406
• ":MEASure:SOURce" on page 444