:EXTernal Trigger Commands 15
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 321
:EXTernal:UNITs
(see page 1138)
Command Syntax
:EXTernal:UNITs <units>
<units> ::= {VOLT | AMPere}
The :EXTernal:UNITs command sets the measurement units for the probe
connected to the external trigger input. Select VOLT for a voltage probe
and select AMPere for a current probe. Measurement results, channel
sensitivity, and trigger level will reflect the measurement units you select.
Query Syntax
:EXTernal:UNITs?
The :CHANnel<n>:UNITs? query returns the current units setting for the
external trigger.
Return Format
<units><NL>
<units> ::= {VOLT | AMP}
See Also "Introduction to :EXTernal Trigger Commands" on page 317
"Introduction to :TRIGger Commands" on page 877
":EXTernal:RANGe" on page 320
":EXTernal:PROBe" on page 319
":CHANnel<n>:UNITs" on page 279