EasyManua.ls Logo

Agilent Technologies InfiniiVision 3000 X-Series

Agilent Technologies InfiniiVision 3000 X-Series
1282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Obsolete and Discontinued Commands 35
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 1071
:MEASure:TMIN
(see page 1138)
Command Syntax
:MEASure:TMIN [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:TMIN command installs a screen measurement and starts an
X-at- Min-Y measurement on the selected waveform. If the optional source
is specified, the current source is modified.
Query Syntax
:MEASure:TMIN? [<source>]
The :MEASure:TMIN? query returns the horizontal axis value at which the
minimum vertical value occurs on the current source. If the optional
source is specified, the current source is modified. If all channels are off,
the query returns 9.9E+37.
Return Format
<value><NL>
<value> ::= time at minimum in NR3 format
See Also "Introduction to :MEASure Commands" on page 406
":MEASure:TMAX" on page 1070
":MEASure:XMAX" on page 467
":MEASure:XMIN" on page 468
NOTE
The :MEASure:TMIN command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:XMIN command (see page 468) instead.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Related product manuals