992 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
32 :WAVeform Commands
:WAVeform:YINCrement
(see page 1138)
Query Syntax
:WAVeform:YINCrement?
The :WAVeform:YINCrement? query returns the y- increment value in volts
for the currently specified source. This value is the voltage difference
between consecutive data values. The y- increment for digital waveforms is
always "1".
Return Format
<value><NL>
<value> ::= y-increment value in the current preamble in 32-bit
floating point NR3 format
See Also • "Introduction to :WAVeform Commands" on page 961
• ":WAVeform:PREamble" on page 976
Example Code • "Example Code" on page 977