:SBUS<n> Commands 27
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 679
:SBUS<n>:FLEXray:TRIGger
(see page 1138)
Command Syntax
:SBUS<n>:FLEXray:TRIGger <condition>
<condition> ::= {FRAMe | ERRor | EVENt}
The :SBUS<n>:FLEXray:TRIGger:TRIGger command sets the FLEXray
trigger on condition:
FRAMe — triggers on specified frames (without errors).
ERRor — triggers on selected active error frames and unknown bus
conditions.
EVENt — triggers on specified FlexRay event/symbol.
Query Syntax
:SBUS<n>:FLEXray:TRIGger?
The :SBUS<n>:FLEXray:TRIGger? query returns the current FLEXray
trigger on condition.
Return Format
<condition><NL>
<condition> ::= {FRAM | ERR | EVEN}
See Also "Introduction to :TRIGger Commands" on page 877
":TRIGger:MODE" on page 886
":SBUS<n>:FLEXray:TRIGger:ERRor:TYPE" on page 680
":SBUS<n>:FLEXray:TRIGger:EVENt:AUToset" on page 681
":SBUS<n>:FLEXray:TRIGger:EVENt:BSS:ID" on page 682
":SBUS<n>:FLEXray:TRIGger:EVENt:TYPE" on page 683
":SBUS<n>:FLEXray:TRIGger:FRAMe:CCBase" on page 684
":SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition" on page 685
":SBUS<n>:FLEXray:TRIGger:FRAMe:ID" on page 686
":SBUS<n>:FLEXray:TRIGger:FRAMe:TYPE" on page 687

Table of Contents