594 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
24 :POWer Commands
:POWer:TRANsient:INEW
(see page 1138)
Command Syntax
:POWer:TRANsient:INEW <value>[suffix]
<value> ::= New current value in NR3 format
[suffix] ::= {A | mA}
The :POWer:TRANsient:INEW command to specify the new load current
value. The new load current will be used as a reference and to trigger the
oscilloscope.
Query Syntax
:POWer:TRANsient:INEW?
The :POWer:TRANsient:INEW query returns the new load current value.
Return Format
<value><NL>
<value> ::= New current value in NR3 format
See Also • ":POWer:TRANsient:APPLy" on page 591
• ":POWer:TRANsient:EXIT" on page 592
• ":POWer:TRANsient:IINitial" on page 593
• ":POWer:TRANsient:NEXT" on page 595

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals