EasyManuals Logo
Home>Agilent Technologies>Test Equipment>InfiniiVision 3000 X-Series

Agilent Technologies InfiniiVision 3000 X-Series Programmer's Guide

Agilent Technologies InfiniiVision 3000 X-Series
1282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #377 background imageLoading...
Page #377 background image
:LISTer Commands 18
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 377
:LISTer:DISPlay
(see page 1138)
Command Syntax
:LISTer:DISPlay <value>
<value> ::= {{OFF | 0} | {SBUS1 | ON | 1} | {SBUS2 | 2} | ALL}
The :LISTer:DISPlay command configures which of the serial buses to
display in the Lister, or whether the Lister is off. "ON" or "1" is the same
as "SBUS1".
When set to "ALL", the decode information for different buses is
interleaved in time.
Serial bus decode must be on before it can be displayed in the Lister.
Query Syntax
:LISTer:DISPlay?
The :LISTer:DISPlay? query returns the Lister display setting.
Return Format
<value><NL>
<value> ::= {OFF | SBUS1 | SBUS2 | ALL}
See Also • "Introduction to :LISTer Commands" on page 375
• ":SBUS<n>:DISPlay" on page 632
• ":LISTer:DATA" on page 376

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals