:LISTer Commands 18
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 377
:LISTer:DISPlay
(see page 1138)
Command Syntax
:LISTer:DISPlay <value>
<value> ::= {{OFF | 0} | {SBUS1 | ON | 1} | {SBUS2 | 2} | ALL}
The :LISTer:DISPlay command configures which of the serial buses to
display in the Lister, or whether the Lister is off. "ON" or "1" is the same
as "SBUS1".
When set to "ALL", the decode information for different buses is
interleaved in time.
Serial bus decode must be on before it can be displayed in the Lister.
Query Syntax
:LISTer:DISPlay?
The :LISTer:DISPlay? query returns the Lister display setting.
Return Format
<value><NL>
<value> ::= {OFF | SBUS1 | SBUS2 | ALL}
See Also "Introduction to :LISTer Commands" on page 375
":SBUS<n>:DISPlay" on page 632
":LISTer:DATA" on page 376