714 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
27 :SBUS<n> Commands
:SBUS<n>:IIC:TRIGger:QUALifier
(see page 1138)
Command Syntax
:SBUS<n>:IIC:TRIGger:QUALifier <value>
<value> ::= {EQUal | NOTequal | LESSthan | GREaterthan}
The :SBUS<n>:IIC:TRIGger:QUALifier command sets the IIC data qualifier
when TRIGger:IIC:TRIGger[:TYPE] is set to READEprom.
Query Syntax
:SBUS<n>:IIC:TRIGger:QUALifier?
The :SBUS<n>:IIC:TRIGger:QUALifier? query returns the current IIC data
qualifier value.
Return Format
<value><NL>
<value> ::= {EQUal | NOTequal | LESSthan | GREaterthan}
See Also • "Introduction to :TRIGger Commands" on page 877
• ":TRIGger:MODE" on page 886
• ":SBUS<n>:IIC:TRIGger[:TYPE]" on page 715