680 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
27 :SBUS<n> Commands
:SBUS<n>:FLEXray:TRIGger:ERRor:TYPE
(see page 1138)
Command Syntax
:SBUS<n>:FLEXray:TRIGger:ERRor:TYPE <error_type>
<error_type> ::= {ALL | HCRC | FCRC}
Selects the FlexRay error type to trigger on. The error type setting is only
valid when the FlexRay trigger mode is set to ERRor.
• ALL — triggers on ALL errors.
• HCRC — triggers on only Header CRC errors.
• FCRC — triggers on only Frame CRC errors.
Query Syntax
:SBUS<n>:FLEXray:TRIGger:ERRor:TYPE?
The :SBUS<n>:FLEXray:TRIGger:ERRor:TYPE? query returns the currently
selected FLEXray error type.
Return Format
<error_type><NL>
<error_type> ::= {ALL | HCRC | FCRC}
See Also • "Introduction to :TRIGger Commands" on page 877
• ":SBUS<n>:FLEXray:TRIGger" on page 679

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals