:SBUS<n> Commands 27
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 773
:SBUS<n>:UART:TRIGger:QUALifier
(see page 1138)
Command Syntax
:SBUS<n>:UART:TRIGger:QUALifier <value>
<value> ::= {EQUal | NOTequal | GREaterthan | LESSthan}
The :SBUS<n>:UART:TRIGger:QUALifier command selects the data qualifier
when :TYPE is set to RDATa, RD1, RD0, RDX, TDATa, TD1, TD0, or TDX
for the trigger when in UART mode.
Query Syntax
:SBUS<n>:UART:TRIGger:QUALifier?
The :SBUS<n>:UART:TRIGger:QUALifier? query returns the current UART
trigger qualifier.
Return Format
<value><NL>
<value> ::= {EQU | NOT | GRE | LESS}
See Also "Introduction to :TRIGger Commands" on page 877
":TRIGger:MODE" on page 886
":SBUS<n>:UART:TRIGger:TYPE" on page 774