896 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
31 :TRIGger Commands
:TRIGger:EBURst Commands
The :TRIGger:EBURst:SOURce command is used to specify the source
channel for the Nth Edge Burst trigger. If an analog channel is selected as
the source, the :TRIGger:EDGE:LEVel command is used to set the Nth Edge
Burst trigger level. If a digital channel is selected as the source, the
:DIGital<n>:THReshold or :POD<n>:THReshold command is used to set the
Nth Edge Burst trigger level.
Table 127 :TRIGger:EBURst Commands Summary
Command Query Options and Query Returns
:TRIGger:EBURst:COUNt
<count> (see
page 897)
:TRIGger:EBURst:COUNt
? (see page 897)
<count> ::= integer in NR1 format
:TRIGger:EBURst:IDLE
<time_value> (see
page 898)
:TRIGger:EBURst:IDLE?
(see page 898)
<time_value> ::= time in seconds
in NR3 format
:TRIGger:EBURst:SLOPe
<slope> (see
page 899)
:TRIGger:EBURst:SLOPe
? (see page 899)
<slope> ::= {NEGative | POSitive}
:TRIGger:EBURst:SOURc
e <source> (see
page 900)
:TRIGger:EBURst:SOURc
e? (see page 900)
<source> ::= {CHANnel<n> |
DIGital<d>}
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format

Table of Contents