1082 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
35 Obsolete and Discontinued Commands
:MTESt:AVERage:COUNt
(see page 1138)
Command Syntax
:MTESt:AVERage:COUNt <count>
<count> ::= an integer from 2 to 65536 in NR1 format
The :MTESt:AVERage:COUNt command sets the number of averages for the
waveforms. With the AVERage acquisition type, the :MTESt:AVERage:COUNt
command specifies the number of data values to be averaged for each time
bucket before the acquisition is considered complete for that time bucket.
Query Syntax
:MTESt:AVERage:COUNt?
The :MTESt:AVERage:COUNt? query returns the currently selected count
value.
Return Format
<count><NL>
<count> ::= an integer from 2 to 65536 in NR1 format
See Also "Introduction to :MTESt Commands" on page 491
":MTESt:AVERage" on page 1081
NOTE
The :MTESt:AVERage:COUNt command is obsolete and is provided for backward
compatibility to previous oscilloscopes. Use the :ACQuire:COUNt command (see page 230)
instead.