922 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
31 :TRIGger Commands
:TRIGger:PATTern:GREaterthan
(see page 1138)
Command Syntax
:TRIGger:PATTern:GREaterthan <greater_than_time>[<suffix>]
<greater_than_time> ::= minimum trigger duration in seconds
in NR3 format
<suffix> ::= {s | ms | us | ns | ps }
The :TRIGger:PATTern:GREaterthan command sets the minimum duration
for the defined pattern when :TRIGger:PATTern:QUALifier is set to
GREaterthan. The command also sets the timeout value when the
:TRIGger:PATTern:QUALifier is set to TIMeout.
Query Syntax
:TRIGger:PATTern:GREaterthan?
The :TRIGger:PATTern:GREaterthan? query returns the minimum duration
time for the defined pattern.
Return Format
<greater_than_time><NL>
See Also • "Introduction to :TRIGger Commands" on page 877
• ":TRIGger:PATTern" on page 919
• ":TRIGger:PATTern:QUALifier" on page 924
• ":TRIGger:MODE" on page 886

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals