770 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
27 :SBUS<n> Commands
:SBUS<n>:UART:TRIGger:BURSt
(see page 1138)
Command Syntax
:SBUS<n>:UART:TRIGger:BURSt <value>
<value> ::= {OFF | 1 to 4096 in NR1 format}
The :SBUS<n>:UART:TRIGger:BURSt command selects the burst value (Nth
frame after idle period) in the range 1 to 4096 or OFF, for the trigger
when in UART mode.
Query Syntax
:SBUS<n>:UART:TRIGger:BURSt?
The :SBUS<n>:UART:TRIGger:BURSt? query returns the current UART
trigger burst value.
Return Format
<value><NL>
<value> ::= {OFF | 1 to 4096 in NR1 format}
See Also "Introduction to :TRIGger Commands" on page 877
":TRIGger:MODE" on page 886
":SBUS<n>:UART:TRIGger:IDLE" on page 772