:FUNCtion Commands 16
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 347
:FUNCtion:OPERation
(see page 1138)
Command Syntax
:FUNCtion:OPERation <operation>
<operation> ::= {ADD | SUBTract | MULTiply | INTegrate | DIFF | FFT
| SQRT | MAGNify | ABSolute | SQUare | LN | LOG | EXP | TEN
| LOWPass | HIGHpass | DIVide | LINear | TRENd | BTIMing | BSTate}
The :FUNCtion:OPERation command sets the desired waveform math
operation:
ADD — Source1 + source2.
SUBTract — Source1 - source2.
MULTiply — Source1 * source2.
INTegrate — Integrate the selected waveform source.
DIFF — Differentiate the selected waveform source.
FFT — Fast Fourier Transform on the selected waveform source.
SQRT — Square root on the selected waveform source.
With the DSOX3ADVMATH advanced math license, these additional
operations are available:
MAGNify — Magnify of the selected waveform source.
ABSolute — Absolute value of the selected waveform source.
SQUare — Square of the selected waveform source.
LN — Natural logarithm on the selected waveform source.
LOG — Common logarithm on the selected waveform source.
EXP — Exponential (e
x
) on the selected waveform source.
TEN — Base 10 exponential (10
x
) on the selected waveform source.
LOWPass — Low-pass filter on the selected waveform source.
HIGHpass — High- pass filter on the selected waveform source.
DIVide — Divide operation on the selected waveform source.
LINear — Ax + B operation on the selected waveform source.
TRENd — Measurement Trend. The math waveform shows measurement
values for each cycle of a selected waveform source.
BTIMing — Chart Logic Bus Timing on the on the selected digital bus.
BSTate — Chart Logic Bus State on the on the selected digital bus.
When the operation is ADD, SUBTract, MULTiply, or DIVide, the
:FUNCtion:SOURce1 and :FUNCtion:SOURce2 commands are used to select
source1 and source2. For all other operations, the :FUNCtion:SOURce1
command selects the waveform source.