868 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
30 :TIMebase Commands
:TIMebase:POSition
(see page 1138)
Command Syntax
:TIMebase:POSition <pos>
<pos> ::= time in seconds from the trigger to the display reference
in NR3 format
The :TIMebase:POSition command sets the time interval between the
trigger event and the display reference point on the screen. The display
reference point is either left, right, or center and is set with the
:TIMebase:REFerence command. The maximum position value depends on
the time/division settings.
Query Syntax
:TIMebase:POSition?
The :TIMebase:POSition? query returns the current time from the trigger to
the display reference in seconds.
Return Format
<pos><NL>
<pos> ::= time in seconds from the trigger to the display reference
in NR3 format
See Also • "Introduction to :TIMebase Commands" on page 866
• ":TIMebase:REFerence" on page 870
• ":TIMebase:RANGe" on page 869
• ":TIMebase:SCALe" on page 871
• ":TIMebase:WINDow:POSition" on page 873
• ":TIMebase:DELay" on page 1093
NOTE
This command is an alias for the :TIMebase:DELay command.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals