900 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
31 :TRIGger Commands
:TRIGger:EBURst:SOURce
(see page 1138)
Command Syntax
:TRIGger:EBURst:SOURce <source>
<source> ::= {CHANnel<n> | DIGital<d>}
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :TRIGger:EBURst:SOURce command selects the input that produces the
Nth edge burst trigger.
Query Syntax
:TRIGger:EBURst:SOURce?
The :TRIGger:EBURst:SOURce? query returns the current Nth edge burst
trigger source. If all channels are off, the query returns "NONE."
Return Format
<source><NL>
<source> ::= {CHAN<n> | DIG<d>}
See Also "Introduction to :TRIGger Commands" on page 877