706 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
27 :SBUS<n> Commands
:SBUS<n>:I2S:WSLow
(see page 1138)
Command Syntax
:SBUS<n>:I2S:WSLow <low_def>
<low_def> ::= {LEFT | RIGHt}
The :SBUS<n>:I2S:WSLow command selects the polarity of the word select
(WS) signal:
LEFT— a word select (WS) state of low indicates left channel data is
active on the I2S bus, and a WS state of high indicates right channel
data is active on the bus.
RIGHt — a word select (WS) state of low indicates right channel data is
active on the I2S bus, and a WS state of high indicates left channel
data is active on the bus.
Query Syntax
:SBUS<n>:I2S:WSLow?
The :SBUS<n>:I2S:WSLow? query returns the currently selected I2S word
select (WS) polarity.
Return Format
<low_def><NL>
<low_def> ::= {LEFT | RIGHt}
See Also "Introduction to :TRIGger Commands" on page 877
":SBUS<n>:I2S:ALIGnment" on page 690
":SBUS<n>:I2S:CLOCk:SLOPe" on page 692
":SBUS<n>:I2S:RWIDth" on page 693
":SBUS<n>:I2S:TWIDth" on page 705