EasyManuals Logo
Home>Agilent Technologies>Test Equipment>InfiniiVision 3000 X-Series

Agilent Technologies InfiniiVision 3000 X-Series Programmer's Guide

Agilent Technologies InfiniiVision 3000 X-Series
1282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #514 background imageLoading...
Page #514 background image
514 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
22 :MTESt Commands
:MTESt:RMODe:TIME
(see page 1138)
Command Syntax
:MTESt:RMODe:TIME <seconds>
<seconds> ::= from 1 to 86400 in NR3 format
When the :MTESt:RMODe command is set to TIME, the
:MTESt:RMODe:TIME command sets the number of seconds for a mask test
to run.
Query Syntax
:MTESt:RMODe:TIME?
The :MTESt:RMODe:TIME? query returns the number of seconds currently
set.
Return Format
<seconds><NL>
<seconds> ::= from 1 to 86400 in NR3 format
See Also • "Introduction to :MTESt Commands" on page 491
• ":MTESt:RMODe" on page 508

Table of Contents