:SBUS<n> Commands 27
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 765
:SBUS<n>:UART:PARity
(see page 1138)
Command Syntax
:SBUS<n>:UART:PARity <parity>
<parity> ::= {EVEN | ODD | NONE}
The :SBUS<n>:UART:PARity command selects the parity to be used with
each message "byte" for the serial decoder and/or trigger when in UART
mode.
Query Syntax
:SBUS<n>:UART:PARity?
The :SBUS<n>:UART:PARity? query returns the current UART parity
setting.
Return Format
<parity><NL>
<parity> ::= {EVEN | ODD | NONE}
See Also • "Introduction to :TRIGger Commands" on page 877
• ":TRIGger:MODE" on page 886
• ":SBUS<n>:UART:TRIGger:TYPE" on page 774