EasyManua.ls Logo

Agilent Technologies InfiniiVision 3000 X-Series - Waveform:source

Agilent Technologies InfiniiVision 3000 X-Series
1282 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:WAVeform Commands 32
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 981
:WAVeform:SOURce
(see page 1138)
Command Syntax
:WAVeform:SOURce <source>
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r> | SBUS{1 | 2}}
for DSO models
<source> ::= {CHANnel<n> | POD{1 | 2} | BUS{1 | 2} | FUNCtion
| MATH | WMEMory<r> | SBUS{1 | 2}}
for MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= {1 | 2}
The :WAVeform:SOURce command selects the analog channel, function,
digital pod, digital bus, reference waveform, or serial decode bus to be
used as the source for the :WAVeform commands.
Function capabilities include add, subtract, multiply, integrate,
differentiate, and FFT (Fast Fourier Transform) operations.
When the :WAVeform:SOURce is the serial decode bus (SBUS1 or SBUS2),
ASCii is the only waveform format allowed, and the :WAVeform:DATA?
query returns a string with timestamps and associated bus decode
information.
With MSO oscilloscope models, you can choose a POD or BUS as the
waveform source. There are some differences between POD and BUS when
formatting and getting data from the oscilloscope:
When POD1 or POD2 is selected as the waveform source, you can
choose the BYTE, WORD, or ASCii formats (see ":WAVeform:FORMat" on
page 971).
When the WORD format is chosen, every other data byte will be 0. The
setting of :WAVeform:BYTeorder controls which byte is 0.
When the ASCii format is chosen, the :WAVeform:DATA? query returns a
string with unsigned decimal values separated by commas.
When BUS1 or BUS2 is selected as the waveform source, you can
choose the WORD or ASCii formats (but not BYTE because bus values
are always returned as 16- bit values).
When the ASCii format is chosen, the :WAVeform:DATA? query returns a
string with hexadecimal bus values, for example: 0x1938,0xff38,...
Query Syntax
:WAVeform:SOURce?
The :WAVeform:SOURce? query returns the currently selected source for
the WAVeform commands.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Related product manuals