:TRIGger Commands 31
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 891
:TRIGger:DELay:ARM:SOURce
(see page 1138)
Command Syntax
:TRIGger:DELay:ARM:SOURce <source>
<source> ::= {CHANnel<n> | DIGital<d>}
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :TRIGger:DELay:ARM:SOURce command selects the input used for the
arming edge in the Edge Then Edge trigger.
Query Syntax
:TRIGger:DELay:ARM:SOURce?
The :TRIGger:DELay:ARM:SOURce? query returns the current arming edge
source.
Return Format
<source><NL>
<source> ::= {CHAN<n> | DIG<d>}
See Also • "Introduction to :TRIGger Commands" on page 877
• ":TRIGger:DELay:ARM:SLOPe" on page 890
• ":TRIGger:DELay:TDELay:TIME" on page 892