:WAVeform Commands 32
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 979
:WAVeform:SEGMented:COUNt
(see page 1138)
Query Syntax
:WAVeform:SEGMented:COUNt?
The :WAVeform:SEGMented:COUNt query returns the number of memory
segments in the acquired data. You can use the
:WAVeform:SEGMented:COUNt? query while segments are being acquired
(although :DIGitize blocks subsequent queries until the full segmented
acquisition is complete).
The segmented memory acquisition mode is enabled with the
:ACQuire:MODE command. The number of segments to acquire is set using
the :ACQuire:SEGMented:COUNt command, and data is acquired using the
:DIGitize, :SINGle, or :RUN commands.
Return Format
<count> ::= an integer from 2 to 1000 in NR1 format (count set by
:ACQuire:SEGMented:COUNt).
See Also • ":ACQuire:MODE" on page 231
• ":ACQuire:SEGMented:COUNt" on page 234
• ":DIGitize" on page 199
• ":SINGle" on page 221
• ":RUN" on page 219
• "Introduction to :WAVeform Commands" on page 961
Example Code • "Example Code" on page 235
NOTE
This command is available when the segmented memory option (Option SGM) is enabled.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?