:POWer Commands 24
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 537
:POWer:HARMonics:APPLy
(see page 1138)
Command Syntax
:POWer:HARMonics:APPLy
The :POWer:HARMonics:APPLy command applies the current harmonics
analysis.
Switching power supplies draw a range of harmonics from the AC mains.
Standard limits are set for these harmonics because these harmonics can
travel back to the supply grid and cause problems with other devices on
the grid.
Use the Current Harmonics analysis to test a switching power supply's
current harmonics to pre- compliance standard of IEC61000- 3- 2 (Class A,
B, C, or D). The analysis presents up to 40 harmonics.
See Also • ":POWer:HARMonics:DATA" on page 538
• ":POWer:HARMonics:DISPlay" on page 539
• ":POWer:HARMonics:FAILcount" on page 540
• ":POWer:HARMonics:LINE" on page 541
• ":POWer:HARMonics:POWerfactor" on page 542
• ":POWer:HARMonics:STANdard" on page 544
• ":POWer:HARMonics:STATus" on page 545
• ":POWer:HARMonics:RUNCount" on page 543
• ":POWer:HARMonics:THD" on page 546