660 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
27 :SBUS<n> Commands
:SBUS<n>:CAN:SIGNal:BAUDrate
(see page 1138)
Command Syntax
:SBUS<n>:CAN:SIGNal:BAUDrate <baudrate>
<baudrate> ::= integer from 10000 to 4000000 in 100 b/s increments,
or 5000000
The :SBUS<n>:CAN:SIGNal:BAUDrate command sets the standard baud rate
of the CAN signal from 10 kb/s to 4 Mb/s in 100 b/s increments. If you
enter a baud rate that is not divisible by 100 b/s, the baud rate is set to
the nearest baud rate divisible by 100 b/s.
You can also set the baud rate of the CAN signal to 5 Mb/s. Fractional
baud rates between 4 Mb/s and 5 Mb/s are not allowed.
If the baud rate you select does not match the system baud rate, false
triggers may occur.
Query Syntax
:SBUS<n>:CAN:SIGNal:BAUDrate?
The :SBUS<n>:CAN:SIGNal:BAUDrate? query returns the current CAN baud
rate setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 10000 to 4000000 in 100 b/s increments,
or 5000000
See Also "Introduction to :TRIGger Commands" on page 877
":SBUS<n>:MODE" on page 633
":SBUS<n>:CAN:TRIGger" on page 663
":SBUS<n>:CAN:SIGNal:DEFinition" on page 661
":SBUS<n>:CAN:SOURce" on page 662