578 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
24 :POWer Commands
:POWer:SIGNals:VSTeady:ONOFf:OFF
(see page 1138)
Command Syntax
:POWer:SIGNals:VSTeady:ONOFf:OFF <value>[suffix]
<value> ::= Expected steady state output Voltage value in NR3 format
[suffix] ::= {V | mV}
The :POWer:SIGNals:VSTeady:ONOFf:OFF command specifies the expected
steady state output DC voltage of the power supply for turn off analysis.
Query Syntax
:POWer:SIGNals:VSTeady:ONOFf:OFF?
The :POWer:SIGNals:VSTeady:ONOFf:OFF query returns the expected steady
state voltage setting.
Return Format
<value><NL>
<value> ::= Expected steady state output Voltage value in NR3 format
See Also ":POWer:ONOFf:APPLy" on page 553
":POWer:SIGNals:AUTosetup" on page 564
":POWer:SIGNals:DURation:ONOFf:OFF" on page 569
":POWer:SIGNals:IEXPected" on page 573
":POWer:SIGNals:OVERshoot" on page 574
":POWer:SIGNals:VMAXimum:ONOFf:OFF" on page 576
":POWer:SIGNals:SOURce:CURRent<i>" on page 581
":POWer:SIGNals:SOURce:VOLTage<i>" on page 582