:CHANnel<n> Commands 10
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 279
:CHANnel<n>:UNITs
(see page 1138)
Command Syntax
:CHANnel<n>:UNITs <units>
<units> ::= {VOLT | AMPere}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:UNITs command sets the measurement units for the
connected probe. Select VOLT for a voltage probe and select AMPere for a
current probe. Measurement results, channel sensitivity, and trigger level
will reflect the measurement units you select.
Query Syntax
:CHANnel<n>:UNITs?
The :CHANnel<n>:UNITs? query returns the current units setting for the
specified channel.
Return Format
<units><NL>
<units> ::= {VOLT | AMP}
See Also • "Introduction to :CHANnel<n> Commands" on page 262
• ":CHANnel<n>:RANGe" on page 277
• ":CHANnel<n>:PROBe" on page 271
• ":EXTernal:UNITs" on page 321