:MEASure Commands 20
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 467
:MEASure:XMAX
(see page 1138)
Command Syntax
:MEASure:XMAX [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1-2 or 1-4 (# of analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:XMAX command installs a screen measurement and starts
an X-at- Max-Y measurement on the selected window. If the optional
source parameter is specified, the current source is modified.
Query Syntax
:MEASure:XMAX? [<source>]
The :MEASure:XMAX? query measures and returns the horizontal axis
value at which the maximum vertical value occurs. If the optional source
is specified, the current source is modified. If all channels are off, the
query returns 9.9E+37.
Return Format
<value><NL>
<value> ::= horizontal value of the maximum in NR3 format
See Also • "Introduction to :MEASure Commands" on page 406
• ":MEASure:XMIN" on page 468
• ":MEASure:TMAX" on page 1070
NOTE
:MEASure:XMAX is an alias for :MEASure:TMAX.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals