:SBUS<n> Commands 27
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 693
:SBUS<n>:I2S:RWIDth
(see page 1138)
Command Syntax
:SBUS<n>:I2S:RWIDth <receiver>
<receiver> ::= 4-32 in NR1 format
The :SBUS<n>:I2S:RWIDth command sets the width of the receiver
(decoded) data word in I2S anywhere from 4 bits to 32 bits.
Query Syntax
:SBUS<n>:I2S:RWIDth?
The :SBUS<n>:I2S:RWIDth? query returns the currently set I2S receiver
data word width.
Return Format
<receiver><NL>
<receiver> ::= 4-32 in NR1 format
See Also "Introduction to :TRIGger Commands" on page 877
":SBUS<n>:I2S:ALIGnment" on page 690
":SBUS<n>:I2S:CLOCk:SLOPe" on page 692
":SBUS<n>:I2S:TWIDth" on page 705
":SBUS<n>:I2S:WSLow" on page 706