:TRIGger Commands 31
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 945
:TRIGger:TV:MODE
(see page 1138)
Command Syntax
:TRIGger:TV:MODE <mode>
<mode> ::= {FIEld1 | FIEld2 | AFIelds | ALINes | LINE | LFIeld1
| LFIeld2 | LALTernate}
The :TRIGger:TV:MODE command selects the TV trigger mode and field.
The LALTernate parameter is not available when :TRIGger:TV:STANdard is
GENeric.
Old forms for <mode> are accepted:
Query Syntax :TRIGger:TV:MODE?
The :TRIGger:TV:MODE? query returns the TV trigger mode.
Return Format
<value><NL>
<value> ::= {FIE1 | FIE2 | AFI | ALIN | LINE | LFI1 | LFI2 | LALT}
See Also "Introduction to :TRIGger Commands" on page 877
":TRIGger:TV:STANdard" on page 948
":TRIGger:MODE" on page 886
<mode> Old Forms Accepted
FIEld1 F1
FIEld2 F2
AFIelds ALLFields, ALLFLDS
ALINes ALLLines
LFIeld1 LINEF1, LINEFIELD1
LFIeld2 LINEF2, LINEFIELD2
LALTernate LINEAlt