650 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
27 :SBUS<n> Commands
:SBUS<n>:A429:TRIGger:TYPE
(see page 1138)
Command Syntax
:SBUS<n>:A429:TRIGger:TYPE <condition>
<condition> ::= {WSTArt | WSTOp | LABel | LBITs | PERRor | WERRor
| GERRor | WGERrors | ALLerrors | LRANge | ABITs
| AOBits | AZBits}
The :SBUS<n>:A429:TRIGger command sets the ARINC 429 trigger on
condition:
• WSTArt — triggers on the start of a word.
• WSTOp — triggers at the end of a word.
• LABel — triggers on the specified label value.
• LBITs — triggers on the label and the other word fields as specified.
• LRANge — triggers on a label within a min/max range.
• PERRor — triggers on words with a parity error.
• WERRor — triggers on an intra-word coding error.
• GERRor — triggers on an inter-word gap error.
• WGERrors — triggers on either a Word or Gap Error.
• ALLerrors — triggers on any of the above errors.
• ABITs — triggers on any bit, which will therefore form an eye diagram.
• AZBits — triggers on any bit with a value of zero.
• AOBits — triggers on any bit with a value of one.
Query Syntax
:SBUS<n>:A429:TRIGger:TYPE?
The :SBUS<n>:A429:TRIGger:TYPE? query returns the current ARINC 429
trigger on condition.
Return Format
<condition><NL>
<condition> ::= {WSTA | WSTO | LAB | LBIT | PERR | WERR | GERR | WGER
| ALL | LRAN | ABIT | AOB | AZB}
Errors • "- 241, Hardware missing" on page 1099
See Also • "Introduction to :SBUS<n> Commands" on page 629
• ":SBUS<n>:MODE" on page 633
• ":SBUS<n>:A429:TRIGger:LABel" on page 645
• ":SBUS<n>:A429:TRIGger:PATTern:DATA" on page 646
• ":SBUS<n>:A429:TRIGger:PATTern:SDI" on page 647
• ":SBUS<n>:A429:TRIGger:PATTern:SSM" on page 648
• ":SBUS<n>:A429:TRIGger:RANGe" on page 649