1060 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
35 Obsolete and Discontinued Commands
:FUNCtion:VIEW
(see page 1138)
Command Syntax
:FUNCtion:VIEW <view>
<view> ::= {{1 | ON} | (0 | OFF}}
The :FUNCtion:VIEW command turns the selected function on or off. When
ON is selected, the function performs as specified using the other
FUNCtion commands. When OFF is selected, function is neither calculated
nor displayed.
Query Syntax
:FUNCtion:VIEW?
The :FUNCtion:VIEW? query returns the current state of the selected
function.
Return Format
<view><NL>
<view> ::= {1 | 0}
NOTE
The :FUNCtion:VIEW command is provided for backward compatibility to previous
oscilloscopes. Use the :FUNCtion:DISPlay command (see page 333) instead.