694 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
27 :SBUS<n> Commands
:SBUS<n>:I2S:SOURce:CLOCk
(see page 1138)
Command Syntax
:SBUS<n>:I2S:SOURce:CLOCk <source>
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital<d>} for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :SBUS<n>:I2S:SOURce:CLOCk controls which signal is used as the
serial clock (SCLK) source by the serial decoder and/or trigger when in
I2S mode.
Query Syntax
:SBUS<n>:I2S:SOURce:CLOCk?
The :SBUS<n>:I2S:SOURce:CLOCk? query returns the current source for the
I2S serial clock (SCLK).
Return Format
<source><NL>
See Also "Introduction to :TRIGger Commands" on page 877
":SBUS<n>:I2S:SOURce:DATA" on page 695
":SBUS<n>:I2S:SOURce:WSELect" on page 696