:SBUS<n> Commands 27
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 769
:SBUS<n>:UART:TRIGger:BASE
(see page 1138)
Command Syntax
:SBUS<n>:UART:TRIGger:BASE <base>
<base> ::= {ASCii | HEX}
The :SBUS<n>:UART:TRIGger:BASE command sets the front panel
UART/RS232 trigger setup data selection option:
ASCii — front panel data selection is from ASCII values.
HEX — front panel data selection is from hexadecimal values.
The :SBUS<n>:UART:TRIGger:BASE setting does not affect the
:SBUS<n>:UART:TRIGger:DATA command which can always set data values
using ASCII or hexadecimal values.
Query Syntax
:SBUS<n>:UART:TRIGger:BASE?
The :SBUS<n>:UART:TRIGger:BASE? query returns the current UART base
setting.
Return Format
<base><NL>
<base> ::= {ASC | HEX}
See Also "Introduction to :TRIGger Commands" on page 877
":TRIGger:MODE" on page 886
":SBUS<n>:UART:TRIGger:DATA" on page 771
NOTE
The :SBUS<n>:UART:TRIGger:BASE command is independent of the
:SBUS<n>:UART:BASE command which affects decode and Lister only.