1078 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
35 Obsolete and Discontinued Commands
:MEASure:VSTArt
(see page 1138)
Command Syntax
:MEASure:VSTArt <vstart_argument>
<vstart_argument> ::= value for vertical marker 1
The :MEASure:VSTArt command moves the vertical marker (Y1 cursor) to
the specified value corresponding to the selected source. The source can
be selected by the MARKer:X1Y1source command.
Query Syntax
:MEASure:VSTArt?
The :MEASure:VSTArt? query returns the current value of the Y1 cursor.
Return Format
<value><NL>
<value> ::= voltage at voltage marker 1 in NR3 format
See Also "Introduction to :MARKer Commands" on page 380
"Introduction to :MEASure Commands" on page 406
":MARKer:Y1Position" on page 389
":MARKer:Y2Position" on page 390
":MARKer:YDELta" on page 391
":MARKer:X1Y1source" on page 383
":MEASure:SOURce" on page 444
":MEASure:TDELta" on page 1068
":MEASure:TSTArt" on page 1072
NOTE
The short form of this command, VSTA, does not follow the defined Long Form to Short
Form Truncation Rules (see page 1140). The normal short form, VST, would be the same for
both VSTArt and VSTOp, so sending VST for the VSTArt command produces an error.
NOTE
The :MEASure:VSTArt command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:Y1Position command (see page 389) instead.