:SBUS<n> Commands 27
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 643
:SBUS<n>:A429:SOURce
(see page 1138)
Command Syntax
:SBUS<n>:A429:SOURce <source>
<source> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
The :SBUS<n>:A429:SOURce command sets the source of the ARINC 429
signal.
Query Syntax
:SBUS<n>:A429:SOURce?
The :SBUS<n>:A429:SOURce? query returns the currently set source of the
ARINC 429 signal.
Use the :TRIGger:LEVel:HIGH and :TRIGger:LEVel:LOW commands to set
the thresold levels for the selected source.
Return Format
<source><NL>
See Also • ":TRIGger:LEVel:HIGH" on page 884
• ":TRIGger:LEVel:LOW" on page 885
• ":TRIGger:MODE" on page 886
• ":SBUS<n>:MODE" on page 633
• ":SBUS<n>:A429:TRIGger:TYPE" on page 650
• ":SBUS<n>:A429:SIGNal" on page 642
• ":SBUS<n>:A429:SPEed" on page 644
• ":SBUS<n>:A429:FORMat" on page 641
• "Introduction to :TRIGger Commands" on page 877