:MTESt Commands 22
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 497
:MTESt:AMASk:UNITs
(see page 1138)
Command Syntax
:MTESt:AMASk:UNITs <units>
<units> ::= {CURRent | DIVisions}
The :MTESt:AMASk:UNITs command alters the way the mask test
subsystem interprets the tolerance parameters for automasking as defined
by :MTESt:AMASk:XDELta and :MTESt:AMASk:YDELta commands.
• CURRent — the mask test subsystem uses the units as set by the
:CHANnel<n>:UNITs command, usually time for
ΔX and voltage for ΔY.
• DIVisions — the mask test subsystem uses the graticule as the
measurement system, so tolerance settings are specified as parts of a
screen division. The mask test subsystem maintains separate XDELta
and YDELta settings for CURRent and DIVisions. Thus, XDELta and
YDELta are not converted to new values when the UNITs setting is
changed.
Query Syntax
:MTESt:AMASk:UNITs?
The :MTESt:AMASk:UNITs query returns the current measurement units
setting for the mask test automask feature.
Return Format
<units><NL>
<units> ::= {CURR | DIV}
See Also • "Introduction to :MTESt Commands" on page 491
• ":MTESt:AMASk:XDELta" on page 498
• ":MTESt:AMASk:YDELta" on page 499
• ":CHANnel<n>:UNITs" on page 279
• ":MTESt:AMASk:SOURce" on page 496
• ":MTESt:SOURce" on page 521
Example Code • "Example Code" on page 491